1 PublicationComparing Wire-Mesh sensor with neutron radiography for measurement of liquid fraction in foam
Ziauddin, M.; Schleicher, E.; Trtik, P.; Knüpfer, L.; Skrypnik, A.; Lappan, T.; Eckert, K.; Heitkam, S.
Abstract
The liquid fraction of foam is an important quantity in engineering process control and essential to interpret foam rheology.
Established measurement tools for the liquid fraction of foam, such as optical measurement or radiography techniques
as well as weighing the foam, are mostly laboratory-based, whereas conductivity-based measurements are limited to the
global measurement without detailed spatial information of liquid fraction. In this work, which combines both types of
measurement techniques, the conductivity-based wire-mesh sensor is compared with neutron radiography. We found a
linear dependency between the liquid fraction of the foam and the wire-mesh readings with a statistical deviation less than
15%. However, the wire-mesh sensor systematically overestimates the liquid fraction which we attribute to liquid bridge
formation between the wires.
Keywords: foam; Wire-mesh-sensor; Neutron radiography
-
Journal of Physics: Condensed Matter 51(2023)1, 015101
Online First (2022) DOI: 10.1088/1361-648X/ac9d16